Place the injection site flat and free of stress, fill the set with water without trapped air and apply 20 kPa above atmospheric pressure. Puncture the intended area with a hypodermic needle of 0,8 mm outside diameter conforming to ISO 7864, hold it for 15 s, remove it and dry the site at once, then watch for leakage for 1 min. For an alternative site design, inject according to the manufacturer's instructions.