ISO 8536-4:2019
Annex A (normative): Physical tests – ISO 8536-4:2019

ISO 8536-4:2019 A.4: Infusion set tensile strength

What does A.4 require?

Pull the infusion set with a steady 15 N force in the direction of its long axis for 15 s and check that it resists that force.

Maintained by Gerard Blokdyk

Other controls in Annex A (normative): Physical tests – ISO 8536-4:2019

Query this from an agent

The graph holds this control, the 0 it maps to, and the evidence behind each claim, over MCP and REST.