ISO 8536-4:2019
Annex A (normative): Physical tests – ISO 8536-4:2019

ISO 8536-4:2019 A.2: Particles in infusion sets

What does A.2 require?

Filter water through a 0,2 µm membrane. Before testing, clean every piece of equipment with that water, wear non-powdered gloves and use a 0,45 µm vacuum filter. In a class N5 clean-air workstation compliant with ISO 14644-1, pass 500 ml of that water through 10 infusion sets that are ready for use, vacuum filter the total collected volume, and count particles on the membrane at ×50 magnification for 25 µm to 50 µm, 51 µm to 100 µm and above 100 µm. Test at least 10 single sets, establish blank counts by testing 10 separate 500 ml portions of the same water, clean again and repeat the background determination if Nb is above 9, then calculate Na and Nb using the coefficients 0,1, 0,2 and 5 so that N = Na - Nb stays at or below 90.