Prepare a bag or container that holds at least 1 000 ml of a 9 g/l sodium chloride solution; cut rigid or semi-rigid containers open at the top, fill the drip chamber to about ⅔, and keep (1 000 ± 5) mm between the drip-chamber liquid level and the set outlet, with a difference of no more than 300 mm from the container level. At maximum flow, record either the time needed to pass 1 000 ml, which must be 10 min or less, or the volume passed in 10 min, which must be at least 1 000 ml. For an integral air-inlet device, shut the air entry for Q0 and then use an unvented rigid container with the closure open for Q1; for a stand-alone device, obtain Q0 without it and Q1 with it fitted after the piercing device, and in both cases require Q1 to be at least Q0 × 0,8.