Bring the complete system to the test temperature, connect the air supply, close every other opening, and hold the inside at 50 kPa above atmospheric pressure for 15 s, then immerse the set in water at (40 ± 1) °C and look for escaping air. Fill the set with degassed distilled water, hold the inside at 20 kPa below atmospheric pressure for 15 s using the same reference, and check whether air enters the set.