ISO 15223-1:2021

International
vISO 15223-1:2021 (fourth edition, July 2021)
2 domains
18 controls

ISO 15223-1 labelling duties for medical devices: when to use the standard symbols for essential information, how to size them and write dates, and what must sit beside each symbol for manufacturer and operator identity, dates and identifiers, sterility and barriers, storage limits, hazards, IVD and infusion values, patient data, translation, repackaging and UDI.

Verified

ISO 15223-1:2021 is a compliance framework from International with 2 domains and 18 controls. The largest domains are Clause 5 and Table 1: Symbols and their requirements of use – ISO 15223-1:2021 (14 controls), Clause 4: General requirements for symbol use – ISO 15223-1:2021 (4 controls). Every control below carries what it requires and what an assessor expects to see.

Maintained by Gerard Blokdyk

Framework summaries on this platform are interpretations grounded in the published source, for educational and compliance planning purposes. They do not reproduce or replace the official standards. Refer to the authoritative source for the definitive text. Framework names and trademarks belong to their respective organisations.

Framework Domains (2)

Clause 4: General requirements for symbol use – ISO 15223-1:2021

4 controls
Controls in the Clause 4: General requirements for symbol use – ISO 15223-1:2021 domain of ISO 15223-1:2021: 4 controls
CodeTitle
iso-15223-1-2021::4.1Validate any symbol proposed for inclusion and show it applies across devices and regions
iso-15223-1-2021::4.2aConvey information essential for proper use with the matching Table 1 symbol
iso-15223-1-2021::4.2bSize each symbol so it stays legible for its purpose
iso-15223-1-2021::4.2cWrite every date and time next to a symbol in the ISO 8601 form

Clause 5 and Table 1: Symbols and their requirements of use – ISO 15223-1:2021

14 controls
Controls in the Clause 5 and Table 1: Symbols and their requirements of use – ISO 15223-1:2021 domain of ISO 15223-1:2021: 14 controls
CodeTitle
iso-15223-1-2021::5.1aManufacturer and authorised representative: place the name and address beside the symbol
iso-15223-1-2021::5.1bDate of manufacture and use-by date: place the date beside the symbol in ISO 8601 form
iso-15223-1-2021::5.1cBatch code, catalogue number, serial number and model number: place the value beside the symbol
iso-15223-1-2021::5.1dImporter and distributor: place the name and address beside the symbol
iso-15223-1-2021::5.1eCountry of manufacture: replace the placeholder with the ISO 3166-1 code
iso-15223-1-2021::5.2aSterility: use the symbol for the actual sterilisation method and identify what is sterile
iso-15223-1-2021::5.2bSterile barrier system symbols: place them with the sterility symbol
iso-15223-1-2021::5.3Storage conditions: write the limit values beside the limit lines
iso-15223-1-2021::5.4Safe use: mark hazards and contents with the matching symbols
iso-15223-1-2021::5.5IVD devices: use the IVD symbols and state the number of tests and the evaluation restriction
iso-15223-1-2021::5.6Transfusion and infusion devices: replace example values with the device's own
iso-15223-1-2021::5.7aPatient and health care information: place the data beside the symbol when used
iso-15223-1-2021::5.7bTranslation and repackaging: name the responsible entity and use the symbols only when another party acted
iso-15223-1-2021::5.7cUnique device identifier: place the UDI symbol beside the UDI carrier if used

What is ISO 15223-1:2021 and who does it apply to?

ISO 15223-1:2021 is a compliance framework from International with 2 domains and 18 controls. ISO 15223-1 labelling duties for medical devices: when to use the standard symbols for essential information, how to size them and write dates, and what must sit beside each symbol for manufacturer and operator identity, dates and identifiers, sterility and barriers, storage limits, hazards, IVD and infusion values, patient data, translation, repackaging and UDI. It is used by organisations to establish and maintain compliance with industry standards and regulatory requirements.

What does ISO 15223-1:2021 actually require?

ISO 15223-1:2021 has 18 controls organised across 2 domains. The largest domains are Clause 5 and Table 1: Symbols and their requirements of use – ISO 15223-1:2021 (14 controls), Clause 4: General requirements for symbol use – ISO 15223-1:2021 (4 controls). Each control defines specific requirements that organisations must implement to achieve compliance.

If I already comply with another framework, how much of ISO 15223-1:2021 do I already cover?

ISO 15223-1:2021 does not currently have cross-framework mappings in our system. Check back as we continuously expand our mapping database.

How do I implement ISO 15223-1:2021?

Start your ISO 15223-1:2021 compliance journey by running a self-assessment on our platform to identify your current compliance posture. Our AI advisory can answer specific questions about ISO 15223-1:2021 requirements, and cross-framework mapping helps you leverage existing controls from other frameworks you may already comply with. Create a free account to access all 18 controls and track your progress.

Start Your Compliance Journey

Create a free account to run self-assessments, get AI advisory, and track your compliance progress across 1,248 frameworks.

Get Started Free →

Free forever, no credit card required