Australia National Pathology Standards (First Edition 2027) 10.06: Frequency of quality control testing: test control samples at a frequency based on the stability and robustness of the method and the risk of patient harm, and additionally when critical consumables change, when confidence in a result's reliability or accuracy is in question, and after substantial equipment maintenance or damage

What does 10.06 require?

How often controls are run is set from how stable and robust the method is and how badly a wrong result could harm a patient, and controls are also run when a critical consumable changes, when a result's reliability or accuracy is doubted, and after substantial maintenance or damage to the equipment.