The ISO general requirements standard for sterile, single-use intravascular catheters: risk and usability, sterility, detectability, biocompatibility, surface, corrosion resistance, tensile force, leakage, hubs, flowrate, power injection, sizing and the information supplied with the catheter, with the annexed test methods. Read against the draft of the 2023 edition.
ISO 10555-1:2023 is a compliance framework from International with 4 domains and 33 controls. The largest domains are Clause 4: Requirements – ISO 10555-1:2023 (18 controls), Normative annexes A to G and I: test methods – ISO 10555-1:2023 (8 controls), Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023 (4 controls). Every control below carries what it requires and what an assessor expects to see.
Framework summaries on this platform are interpretations grounded in the published source, for educational and compliance planning purposes. They do not reproduce or replace the official standards. Refer to the authoritative source for the definitive text. Framework names and trademarks belong to their respective organisations.
Framework Domains (4)
Clause 4: Requirements – ISO 10555-1:2023
| Code | Title |
|---|---|
| iso-10555-1-2023::4.1 | Risk management to ISO 14971 |
| iso-10555-1-2023::4.10 | Leakage under pressurization |
| iso-10555-1-2023::4.11 | Leakage during aspiration |
| iso-10555-1-2023::4.12 | Female hub connector conformity |
| iso-10555-1-2023::4.13 | Gravity flow rate per lumen |
| iso-10555-1-2023::4.14 | Burst pressure during power injection |
| iso-10555-1-2023::4.15 | Packaging system to ISO 11607-1 |
| iso-10555-1-2023::4.16 | Simulated use, kink and torque strength |
| iso-10555-1-2023::4.17 | Coating integrity and particulate testing |
| iso-10555-1-2023::4.18 | Distal tip stiffness for neurovascular use |
| iso-10555-1-2023::4.2 | Usability engineering programme |
| iso-10555-1-2023::4.3 | Validated sterilization and post-sterilization conformity |
| iso-10555-1-2023::4.4 | Shelf life supported by ageing assessment |
| iso-10555-1-2023::4.5 | Visibility under X-ray or by other methods |
| iso-10555-1-2023::4.6 | Biological safety to ISO 10993-1 |
| iso-10555-1-2023::4.7 | Clean external surface and lubricant |
| iso-10555-1-2023::4.8 | Corrosion resistance of metal parts |
| iso-10555-1-2023::4.9 | Peak tensile force of catheter sections |
Clause 5: Designation of nominal size – ISO 10555-1:2023
| Code | Title |
|---|---|
| iso-10555-1-2023::5.1 | Outside diameter and nominal size |
| iso-10555-1-2023::5.2 | Inside diameter and nominal size |
| iso-10555-1-2023::5.3 | Nominal effective length |
Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023
| Code | Title |
|---|---|
| iso-10555-1-2023::6.1 | Information supplied with the catheter |
| iso-10555-1-2023::6.2 | Device and primary packaging marking |
| iso-10555-1-2023::6.3 | Instructions for use content |
| iso-10555-1-2023::6.4 | Secondary packaging marking |
Normative annexes A to G and I: test methods – ISO 10555-1:2023
| Code | Title |
|---|---|
| iso-10555-1-2023::Annex-A | Corrosion resistance test |
| iso-10555-1-2023::Annex-B | Determining the peak tensile force |
| iso-10555-1-2023::Annex-C | Liquid leakage under pressure test |
| iso-10555-1-2023::Annex-D | Air leakage into hub during aspiration |
| iso-10555-1-2023::Annex-E | Determining catheter flowrate |
| iso-10555-1-2023::Annex-F | Static burst pressure test |
| iso-10555-1-2023::Annex-G | Power injection flowrate and pressure test |
| iso-10555-1-2023::Annex-I | Air leakage test under water |
What is ISO 10555-1:2023 and who does it apply to?
ISO 10555-1:2023 is a compliance framework from International with 4 domains and 33 controls. The ISO general requirements standard for sterile, single-use intravascular catheters: risk and usability, sterility, detectability, biocompatibility, surface, corrosion resistance, tensile force, leakage, hubs, flowrate, power injection, sizing and the information supplied with the catheter, with the annexed test methods. Read against the draft of the 2023 edition. It is used by organisations to establish and maintain compliance with industry standards and regulatory requirements.
What does ISO 10555-1:2023 actually require?
ISO 10555-1:2023 has 33 controls organised across 4 domains. The largest domains are Clause 4: Requirements – ISO 10555-1:2023 (18 controls), Normative annexes A to G and I: test methods – ISO 10555-1:2023 (8 controls), Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023 (4 controls). Each control defines specific requirements that organisations must implement to achieve compliance.
If I already comply with another framework, how much of ISO 10555-1:2023 do I already cover?
ISO 10555-1:2023 does not currently have cross-framework mappings in our system. Check back as we continuously expand our mapping database.
How do I implement ISO 10555-1:2023?
Start your ISO 10555-1:2023 compliance journey by running a self-assessment on our platform to identify your current compliance posture. Our AI advisory can answer specific questions about ISO 10555-1:2023 requirements, and cross-framework mapping helps you leverage existing controls from other frameworks you may already comply with. Create a free account to access all 33 controls and track your progress.
Start Your Compliance Journey
Create a free account to run self-assessments, get AI advisory, and track your compliance progress across 1,381 frameworks.
Get Started Free →Free forever, no credit card required