Back to Frameworks

ISO 10555-1:2023

International
vISO 10555-1:2023 (third edition), read against ISO/DIS 10555-1:2022
4 domains
33 controls

The ISO general requirements standard for sterile, single-use intravascular catheters: risk and usability, sterility, detectability, biocompatibility, surface, corrosion resistance, tensile force, leakage, hubs, flowrate, power injection, sizing and the information supplied with the catheter, with the annexed test methods. Read against the draft of the 2023 edition.

Verified

ISO 10555-1:2023 is a compliance framework from International with 4 domains and 33 controls. The largest domains are Clause 4: Requirements – ISO 10555-1:2023 (18 controls), Normative annexes A to G and I: test methods – ISO 10555-1:2023 (8 controls), Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023 (4 controls). Every control below carries what it requires and what an assessor expects to see.

Maintained by Gerard Blokdyk

Framework summaries on this platform are interpretations grounded in the published source, for educational and compliance planning purposes. They do not reproduce or replace the official standards. Refer to the authoritative source for the definitive text. Framework names and trademarks belong to their respective organisations.

Framework Domains (4)

Clause 4: Requirements – ISO 10555-1:2023

18 controls
Controls in the Clause 4: Requirements – ISO 10555-1:2023 domain of ISO 10555-1:2023: 18 controls
CodeTitle
iso-10555-1-2023::4.1Risk management to ISO 14971
iso-10555-1-2023::4.10Leakage under pressurization
iso-10555-1-2023::4.11Leakage during aspiration
iso-10555-1-2023::4.12Female hub connector conformity
iso-10555-1-2023::4.13Gravity flow rate per lumen
iso-10555-1-2023::4.14Burst pressure during power injection
iso-10555-1-2023::4.15Packaging system to ISO 11607-1
iso-10555-1-2023::4.16Simulated use, kink and torque strength
iso-10555-1-2023::4.17Coating integrity and particulate testing
iso-10555-1-2023::4.18Distal tip stiffness for neurovascular use
iso-10555-1-2023::4.2Usability engineering programme
iso-10555-1-2023::4.3Validated sterilization and post-sterilization conformity
iso-10555-1-2023::4.4Shelf life supported by ageing assessment
iso-10555-1-2023::4.5Visibility under X-ray or by other methods
iso-10555-1-2023::4.6Biological safety to ISO 10993-1
iso-10555-1-2023::4.7Clean external surface and lubricant
iso-10555-1-2023::4.8Corrosion resistance of metal parts
iso-10555-1-2023::4.9Peak tensile force of catheter sections

Clause 5: Designation of nominal size – ISO 10555-1:2023

3 controls
Controls in the Clause 5: Designation of nominal size – ISO 10555-1:2023 domain of ISO 10555-1:2023: 3 controls
CodeTitle
iso-10555-1-2023::5.1Outside diameter and nominal size
iso-10555-1-2023::5.2Inside diameter and nominal size
iso-10555-1-2023::5.3Nominal effective length

Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023

4 controls
Controls in the Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023 domain of ISO 10555-1:2023: 4 controls
CodeTitle
iso-10555-1-2023::6.1Information supplied with the catheter
iso-10555-1-2023::6.2Device and primary packaging marking
iso-10555-1-2023::6.3Instructions for use content
iso-10555-1-2023::6.4Secondary packaging marking

Normative annexes A to G and I: test methods – ISO 10555-1:2023

8 controls
Controls in the Normative annexes A to G and I: test methods – ISO 10555-1:2023 domain of ISO 10555-1:2023: 8 controls
CodeTitle
iso-10555-1-2023::Annex-ACorrosion resistance test
iso-10555-1-2023::Annex-BDetermining the peak tensile force
iso-10555-1-2023::Annex-CLiquid leakage under pressure test
iso-10555-1-2023::Annex-DAir leakage into hub during aspiration
iso-10555-1-2023::Annex-EDetermining catheter flowrate
iso-10555-1-2023::Annex-FStatic burst pressure test
iso-10555-1-2023::Annex-GPower injection flowrate and pressure test
iso-10555-1-2023::Annex-IAir leakage test under water

What is ISO 10555-1:2023 and who does it apply to?

ISO 10555-1:2023 is a compliance framework from International with 4 domains and 33 controls. The ISO general requirements standard for sterile, single-use intravascular catheters: risk and usability, sterility, detectability, biocompatibility, surface, corrosion resistance, tensile force, leakage, hubs, flowrate, power injection, sizing and the information supplied with the catheter, with the annexed test methods. Read against the draft of the 2023 edition. It is used by organisations to establish and maintain compliance with industry standards and regulatory requirements.

What does ISO 10555-1:2023 actually require?

ISO 10555-1:2023 has 33 controls organised across 4 domains. The largest domains are Clause 4: Requirements – ISO 10555-1:2023 (18 controls), Normative annexes A to G and I: test methods – ISO 10555-1:2023 (8 controls), Clause 6: Information to be supplied with the catheter – ISO 10555-1:2023 (4 controls). Each control defines specific requirements that organisations must implement to achieve compliance.

If I already comply with another framework, how much of ISO 10555-1:2023 do I already cover?

ISO 10555-1:2023 does not currently have cross-framework mappings in our system. Check back as we continuously expand our mapping database.

How do I implement ISO 10555-1:2023?

Start your ISO 10555-1:2023 compliance journey by running a self-assessment on our platform to identify your current compliance posture. Our AI advisory can answer specific questions about ISO 10555-1:2023 requirements, and cross-framework mapping helps you leverage existing controls from other frameworks you may already comply with. Create a free account to access all 33 controls and track your progress.

Start Your Compliance Journey

Create a free account to run self-assessments, get AI advisory, and track your compliance progress across 1,381 frameworks.

Get Started Free →

Free forever, no credit card required