Each section of the catheter must be tested for peak tensile force under Annex B, on the whole device or as a separate section, and must meet Table 1: 3 N from 0,55 mm to below 0,75 mm outside diameter, 5 N from 0,75 mm to below 1,15 mm, 10 N from 1,15 mm to below 1,85 mm, and 15 N from 1,85 mm upward. For a distal tip shorter than 3 mm and its shaft junction, the manufacturer must set the test method and force from a risk assessment; the same applies where clinical forces are judged higher than Table 1. Tubing below 0,55 mm outside diameter, measured before hydration for a hydratable catheter, falls outside Table 1, and the manufacturer should test it to Annex B and derive the force from a risk assessment.