ISO 10555-1:2023
Clause 4: Requirements – ISO 10555-1:2023

ISO 10555-1:2023 4.7: Clean external surface and lubricant

What does 4.7 require?

When viewed with normal or corrected vision at a magnification of at least x 2,5, the outside of the catheter must show no foreign matter. The effective length, including the distal end, must have no surface defects that could cause emboli or vessel trauma. If a lubricant is used, it must not appear as visible drops on the outer surface under the same viewing conditions.