ISO 10555-1:2023
Normative annexes A to G and I: test methods – ISO 10555-1:2023

ISO 10555-1:2023 Annex-E: Determining catheter flowrate

What does Annex-E require?

A constant-level tank supplies water or another clinically relevant fluid at 22 ± 5 °C. The tank has a hydrostatic head of 1000 ± 5 mm and can deliver 525 ± 25 ml/min free flow, and it connects to the catheter through a male 6 % Luer taper fitting. Effluent is collected for no less than 30 s and quantified by volume or mass to within 1 %, or a flow sensor with 5 % accuracy is used once flow is steady. Three determinations are made on each applicable lumen, the arithmetic average is rounded to the nearest whole millilitre per minute, and the report states the average for each lumen and, across all samples, the highest and lowest values, the average and the standard deviation, along with any deviations and the date.