Corrective action policies and procedures must be available and followed as necessary to keep the laboratory testing patient specimens in a way that ensures accurate and reliable results and reports. The laboratory must document all corrective actions taken, including when test systems do not meet the verified or established performance specifications (equipment or methods operating outside parameters, patient values outside the reportable range, or reference intervals found inappropriate for the population); when control or calibration results fail the acceptability criteria, in which case every patient result in the unacceptable run and since the last acceptable run must be evaluated for adverse effect and the necessary action taken to ensure accurate reporting; and when the reagent and specimen storage criteria under section 493.1252(b) are not met.