Test systems must be selected by the laboratory and testing performed following the manufacturer's instructions in a manner that gives results within the performance specifications the laboratory determined under section 493.1253. The laboratory must define criteria for the conditions essential to proper storage of reagents and specimens, accurate and reliable test system operation and result reporting, consistent with the manufacturer's instructions, and must monitor and document those conditions, which where applicable include water quality, temperature, humidity and protection of equipment from electrical fluctuations and interruptions that could affect results. Reagents, solutions, culture media, control and calibration materials and other supplies must be labelled with identity and, where significant, titer, strength or concentration, storage requirements, preparation and expiration dates and other information needed for proper use; must not be used past their expiration date or when deteriorated or of substandard quality; and components of reagent kits of different lot numbers must not be interchanged unless the manufacturer specifies otherwise.