For unmodified manufacturer's equipment, instruments or test systems the laboratory must perform and document maintenance as defined by the manufacturer at least at the specified frequency, and function checks as defined by the manufacturer at least at the specified frequency, with function checks within the manufacturer's established limits before patient testing is conducted. For equipment, instruments or test systems developed in-house, commercially available but modified by the laboratory, or for which the manufacturer provides no maintenance or function check protocol, the laboratory must establish a maintenance protocol that ensures the performance needed for accurate and reliable results and reporting, perform and document that maintenance, define a function check protocol with the same aim, and perform and document the function checks, including background or baseline checks, which must be within the laboratory's established limits before patient testing.