Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)
Part B: Conditions of certification – Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)

Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1) C29: Annual testing of testable backflow devices

What does C29 require?

Any testable backflow prevention devices fitted (under AS/NZS 3500.1) need testing at least once every 12 months, following AS 2845.3; the tester must be a licensed plumber holding accreditation to test backflow prevention devices. Failures must be fixed and the device retested until it complies. The guideline notes that C68 requires test records.

Maintained by Gerard Blokdyk

Other controls in Part B: Conditions of certification – Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)