If testing shows a cabinet falls short on (a) for Class I, its inward air velocity or the integrity of its HEPA filter, or (b) for Class II, its air barrier containment or the integrity of its exhaust HEPA filter, and the defect has not been fixed, the cabinet must be clearly labelled as defective and not used for dealings that produce GMO-containing aerosols.
The graph holds this control, the 0 it maps to, and the evidence behind each claim, over MCP and REST.