Evaporate 50 ml of S1 to dryness in a tared dish at a temperature just below boiling, then keep it in an oven at 105 °C until the mass stays constant; handle 50 ml of S0 the same way. Report the difference between the two residual masses in milligrams.
The graph holds this control, the 0 it maps to, and the evidence behind each claim, over MCP and REST.