ISO 8536-4:2019
Annex B (normative): Chemical tests – ISO 8536-4:2019

ISO 8536-4:2019 B.1: Making the extract and the blank

What does B.1 require?

Assemble a closed circuit from three infusion sets that have been sterilized plus a boiling flask of borosilicate glass holding 300 ml. Hold the flask contents at (37 ± 1) °C and pump 250 ml of water that meets grade 2 or grade 1 of ISO 3696 around the circuit for 2 h at 1 l/h, driven by a peristaltic pump and the shortest workable length of silicone tubing. Collect this solution and cool it to give S1. Make S0 in the same way but leave out the infusion sets, then use both solutions for the chemical tests.

Maintained by Gerard Blokdyk

Other controls in Annex B (normative): Chemical tests – ISO 8536-4:2019

Query this from an agent

The graph holds this control, the 0 it maps to, and the evidence behind each claim, over MCP and REST.