ISO 14644-1:2015
Annex A (normative): Reference method for classification by particle concentration – ISO 14644-1:2015

ISO 14644-1:2015 A.2: Use a discrete particle counter able to size and count at the required sizes, with a valid calibration

What does A.2 require?

The reference method relies on an instrument that counts particles one by one and can show or log both the count and the size, with discrimination fine enough to capture the total concentration in each size band relevant to the class concerned. The counter must carry a calibration certificate that is in date; how often and how it is calibrated is recommended to follow accepted current practice as laid down in ISO 21501-4, and if a counter cannot be calibrated to all of that standard's tests, the choice to use it anyway is written into the test report.

Maintained by Gerard Blokdyk

Other controls in Annex A (normative): Reference method for classification by particle concentration – ISO 14644-1:2015