Australia Therapeutic Goods (Medical Devices) Regulations 2002
Schedule 1 Part 2: other essential principles (clauses 7 to 15) – Australia Therapeutic Goods (Medical Devices) Regulations 2002

Australia Therapeutic Goods (Medical Devices) Regulations 2002 sch1-12.12-12.13: Essential principles 12.12 and 12.13: devices that deliver energy or substances must set and hold the delivered rate and amount accurately, prevent accidental dangerous release, flag inadequate delivery and label controls clearly; active implants must carry a code readable without surgery identifying type, manufacturer and year

What does sch1-12.12-12.13 require?

A device meant to administer energy or a substance must let the delivered rate and amount be set and maintained accurately for patient and user safety, prevent as far as possible accidental release of dangerous levels, have a means of indicating or, where appropriate, preventing inadequate delivery that could cause danger, show clearly the function of each control and indicator, and present any on-screen instructions or parameters so users (and patients where relevant) can understand them. Every active implant must carry some code or distinctive feature (shown, transmitted or built in) that can be read without operating on the patient and that reveals what type of device it is, who made it and the year it was made.

Maintained by Gerard Blokdyk

Other controls in Schedule 1 Part 2: other essential principles (clauses 7 to 15) – Australia Therapeutic Goods (Medical Devices) Regulations 2002