Where any proposed dealing with GM micro-organisms would produce aerosols carrying Risk Group 2 GM micro-organisms, a biological safety cabinet must be present in the facility, or other equipment designed to contain aerosols that the Regulator has specifically approved in writing. A Class I or Class II cabinet that is installed must be installed to the requirements of AS 2252.4. This sits under the Part A heading for dealings with GM micro-organisms, so it applies in addition to the other requirements when any proposed dealing involves them.