Every biological safety cabinet of Class I or Class II installed and used for GMO procedures must be inspected and tested by a qualified person against the performance requirements in Section 5.2 of AS 2252.1 (Class I) or Section 5.2 of AS 2252.2 (Class II). Testing is needed at least every 12 months and again whenever the cabinet is relocated, has mechanical or electrical maintenance, or has its HEPA filters replaced.