Australia OGTR Guidelines for Certification of a PC2 Invertebrate Facility (Version 3.1)
Part B: Conditions of certification – Australia OGTR Guidelines for Certification of a PC2 Invertebrate Facility (Version 3.1)

Australia OGTR Guidelines for Certification of a PC2 Invertebrate Facility (Version 3.1) B10: Work not allowed in a PC2 invertebrate facility

What does B10 require?

Unless the Regulator agrees in writing, the facility must not be used for: (a) dealings with GMOs that a licence or legislation says need containment above PC2; (b) housing, keeping or rearing terrestrial vertebrates or aquatic organisms beyond the shortest time needed to finish procedures on them; (c) growing plants, unless they form an integral part of work on GM invertebrates or on invertebrates that contain GMOs; (d) work on GM invertebrates, or on non-GM invertebrates carrying GM micro-organisms, that spend part of their life cycle in water, where any single container holds more than 25 litres; (e) GMO cultures where any single container holds more than 25 litres; or (f) anything else the Regulator has prohibited in writing.