IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

International (IAF MLA and Global ACI Arrangement signatories)
vGlobal ACI TECH-3-007 (M) v1.0 (issued 10 July 2026, implemented 28 August 2026), carrying forward IAF MD 9:2023; IAF MD 9:2022 Issue 4 is the earlier edition
6 domains
21 controls

The mandatory accreditation criteria added to ISO/IEC 17021-1 for certifying medical device quality management systems to ISO 13485: regulatory compliance checks, impartiality, auditor and decision maker competence and experience, technical areas, audit time, no sampling of manufacturing sites, scope precision, major nonconformities, vigilance review in surveillance and unannounced audits.

Verified

IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) is a compliance framework from International (IAF MLA and Global ACI Arrangement signatories) with 6 domains and 21 controls. The largest domains are Clause 9: Process requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) (8 controls), Annexes A to D: Technical areas, knowledge and skills, auditor qualification, audit time – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) (4 controls), Clause 7: Resource requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) (4 controls). Every control below carries what it requires and what an assessor expects to see.

Maintained by Gerard Blokdyk

Framework summaries on this platform are interpretations grounded in the published source, for educational and compliance planning purposes. They do not reproduce or replace the official standards. Refer to the authoritative source for the definitive text. Framework names and trademarks belong to their respective organisations.

Framework Domains (6)

Annexes A to D: Technical areas, knowledge and skills, auditor qualification, audit time – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

4 controls
Controls in the Annexes A to D: Technical areas, knowledge and skills, auditor qualification, audit time – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) domain of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485): 4 controls
CodeTitle
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::AUse the Annex A technical areas to define scope, identify technical qualification needs and select the team
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::BDefine the Annex B knowledge and skills per certification function and apply the parts and services test
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::CQualify auditors by education, four years' experience, Annex B competence and continuing development
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::DStart initial audit time from Table D.1 and adjust it within the stated limits

Clause 4: Principles – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

1 controls
Controls in the Clause 4: Principles – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) domain of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485): 1 controls
CodeTitle
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 4.4.1Determine that the client has evaluated regulatory compliance and acted on non-compliance

Clause 5: General requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

2 controls
Controls in the Clause 5: General requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) domain of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485): 2 controls
CodeTitle
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 5.1.2Hold client agreements that allow audit report information to go to regulators recognising ISO 13485
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 5.2.3Keep the body and its auditors clear of device, QMS and representative involvements that compromise impartiality

Clause 7: Resource requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

4 controls
Controls in the Clause 7: Resource requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) domain of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485): 4 controls
CodeTitle
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 7.1.1Treat the scheme as medical devices and applicable law, and meet the Annex B competence requirements
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 7.2.4Demonstrate auditor competence per Annex C and authorise auditors by Annex A technical area
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 7.2.5Require supervised audit experience before first authorisation, more for team leaders
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 7.2.8Ensure the certification decision makers collectively meet Annex B competence

Clause 8: Information requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

2 controls
Controls in the Clause 8: Information requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) domain of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485): 2 controls
CodeTitle
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 8.1.3Give regulators information on certifications granted, suspended or withdrawn where the law or the authority requires
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 8.2.1Document the scope precisely and exclude nothing that affects product safety and quality

Clause 9: Process requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

8 controls
Controls in the Clause 9: Process requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) domain of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485): 8 controls
CodeTitle
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.1.2.1Decide and record whether outsourced processes need specific audit team competence
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.1.4Set audit time from Annex D and the TECH-3-004 rules, with regulatory and dossier time added, and justify it
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.1.5Never sample sites that design, develop or manufacture medical devices
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.2.2.1Field a team competent for the technical area, with sterilisation competence where devices are sterile
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.3.1Avoid repeating audit work already done under a regulatory scheme that covers ISO 13485, and hold stage 1 on site for higher-risk devices
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.4.5Classify as major the nonconformities that need accepted and verified corrective action
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.6.2.2Review adverse event notifications, advisory notices and recalls in surveillance
iaf-md-9-global-aci-tech-3-007-application-of-iso-iec-17021-1-for-medical-device-quality-management-system-certification-iso-13485::MD 9.6.4.2Hold short-notice or unannounced audits when devices, safety information, significant changes or the law call for them

What is IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) and who does it apply to?

IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) is a compliance framework from International (IAF MLA and Global ACI Arrangement signatories) with 6 domains and 21 controls. The mandatory accreditation criteria added to ISO/IEC 17021-1 for certifying medical device quality management systems to ISO 13485: regulatory compliance checks, impartiality, auditor and decision maker competence and experience, technical areas, audit time, no sampling of manufacturing sites, scope precision, major nonconformities, vigilance review in surveillance and unannounced audits. It is used by organisations to establish and maintain compliance with industry standards and regulatory requirements.

What does IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) actually require?

IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) has 21 controls organised across 6 domains. The largest domains are Clause 9: Process requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) (8 controls), Annexes A to D: Technical areas, knowledge and skills, auditor qualification, audit time – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) (4 controls), Clause 7: Resource requirements – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) (4 controls). Each control defines specific requirements that organisations must implement to achieve compliance.

If I already comply with another framework, how much of IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) do I already cover?

IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) does not currently have cross-framework mappings in our system. Check back as we continuously expand our mapping database.

How do I implement IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)?

Start your IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) compliance journey by running a self-assessment on our platform to identify your current compliance posture. Our AI advisory can answer specific questions about IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) requirements, and cross-framework mapping helps you leverage existing controls from other frameworks you may already comply with. Create a free account to access all 21 controls and track your progress.