IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)
Clause 4: Principles – IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485)

IAF MD 9 / Global ACI TECH-3-007 Application of ISO/IEC 17021-1 for Medical Device Quality Management System Certification (ISO 13485) MD 4.4.1: Determine that the client has evaluated regulatory compliance and acted on non-compliance

What does MD 4.4.1 require?

Because ISO 13485 obliges the certified organisation to meet the statutory and regulatory requirements for the safety and performance of its devices, and maintaining and evaluating legal compliance is the client's own responsibility, the certification body's task is to determine that the client has evaluated how it complies with statute and regulation and can show appropriate action where it fell short, including reporting notifiable incidents to the regulatory authority.