ISO 8536-4:2019 7.8: Drip chamber and drip tube dimensions

What does 7.8 require?

The drip chamber must let an observer watch the falling drops without interruption, and liquid enters through a tube that reaches inside the chamber. The gap from the drip tube end to the chamber outlet must be at least 40 mm, the gap from the drip tube to the fluid filter must be at least 20 mm, and the chamber wall must remain at least 5 mm away from the drip tube end. By design, the drip tube must give 1 ± 0,1 ml or 1 ± 0,1 g from 20 drops, or from 60 drops, when distilled water is held at 23 ± 2 °C and delivered at 50 ± 10 drops/min, and the chamber should allow priming and make it easy.