Indicator light colours carry the meanings of Table 2: red for a warning needing immediate operator response, yellow for caution needing prompt response, green for ready for use, any other colour for other meanings, with alarm indicator detail in IEC 60601-1-8 and alphanumeric displays not counted as indicator lights; the colour red is used only for a control that interrupts a function in an emergency; compliance is checked by inspection.
The graph holds this control, the 0 it maps to, and the evidence behind each claim, over MCP and REST.