IEC 60601-1 - Medical Electrical Equipment Safety 7.7: Colours of the insulation of conductors

What does 7.7 require?

A protective earth conductor has green and yellow insulation along its whole length; internal conductors forming protective earth connections are identified green and yellow at least at their terminations; green and yellow is used only for protective earth conductors, those internal connections, potential equalisation conductors and functional earth conductors; neutral conductors in power supply cords are light blue per IEC 60227-1 or 60245-1; and cord conductor colours follow those standards; compliance is checked by inspection.