Switches controlling power, including mains switches, show their on and off positions either with the IEC 60417 on and off symbols or by a nearby indicator light or some other unambiguous method; bistable push buttons carry the stand-by-type symbol and their state is shown by an adjacent light or other unambiguous means, momentary push buttons the momentary symbol or a state indication; control and switch positions are shown with numbers, letters or another visual method, and where changing a setting in normal use could create unacceptable risk to the patient the control is paired with an indicator, or shows which way the setting is changing; numeric indications are in SI units per ISO 80000-1 except the listed permitted units (degree, minute, hour, day, litre, mmHg, cmH2O, bar, electron volt and the like). Compliance is checked by inspection together with the 7.1 tests.