Clause 16: ME systems – IEC 60601-1 - Medical Electrical Equipment Safety

IEC 60601-1 - Medical Electrical Equipment Safety 16.8: Interruption of the power supply to parts of an ME system

What does 16.8 require?

Interruption and restoration of power to the whole system or to any part of it causes only an interruption of the intended function, without basic safety or essential performance being lost; this is tested by breaking and restoring each relevant connection singly and then all of them together.