Where non-ME equipment sits within the patient environment, any of its parts the operator can touch during routine maintenance or calibration after removing covers or connectors without a tool operate at no more than the 8.4.2 voltage limits and draw their supply from a source that two means of operator protection isolate from the mains; checked by inspection.