IEC 60601-1 - Medical Electrical Equipment Safety 11.4: 11.4 ME equipment and ME systems intended for use with flammable anaesthetics

Equipment, systems or parts that the accompanying documents declare suitable for flammable anaesthetics (category AP), or for flammable anaesthetics combined with oxidants (category APG), meet the applicable requirements of the normative Annex G on protection against ignition of flammable anaesthetic mixtures.