Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)
Part A: Requirements for certification – Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)

Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1) R9: Backflow protection of the water supply

What does R9 require?

Wherever a device or system able to contaminate drinking (potable) water is joined, whether directly or by an indirect route, to any section of a water service, it must be protected against backflow by a registered testable device carrying a high hazard rating that guards against back-pressure as well as back-siphonage, as AS/NZS 3500.1 requires. The guideline notes this covers any water supplied to the facility, for example sinks, basins and the supply to an autoclave.