Where another reticulated service, such as air, gas or steam, is connected to primary containment equipment for large scale work (a closed system or LWTS, for example) and could form a cross-connection leading to a GMO being released beyond the certified facility, the need for backflow prevention must be decided by a risk assessment. The guideline suggests that a filter with a pore size of 0.2 µm or less generally suits air and gas, and that suitable filters or other mechanisms that stop backflow should be considered for steam lines.