Part A: Requirements for certification – Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)

Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1) R24: Secondary containment for the closed system

What does R24 require?

There must be secondary containment, bunding for example, to hold any leak from the closed system. Its capacity must cover (a) the fluid volume of the largest single container, or of a group of containers where interconnection could let several leak at once, plus (b) whatever volume of disinfectant may be used, with extra capacity so that expected general fluid movement does not breach it.

Maintained by Gerard Blokdyk

Other controls in Part A: Requirements for certification – Australia OGTR Guidelines for Certification of a PC2 Large Scale Facility (Version 3.1)