Australia OGTR Guidelines for Certification of a PC2 Laboratory (Version 3.2) B21: Cabinet that fails its test

What does B21 require?

If a test finds a cabinet falls short on inward air velocity or on HEPA filter integrity (Class I), or on air barrier containment or the integrity of the exhaust HEPA filter (Class II), and the fault has not been fixed, the cabinet must be clearly marked as defective and not used for procedures that generate aerosols containing GMOs.