Australia OGTR Guidelines for Certification of a PC2 Laboratory (Version 3.2) B19: Testing biological safety cabinets

What does B19 require?

Every biological safety cabinet of Class I or Class II used for GMO procedures must be inspected and tested by a qualified person against the performance requirements in Section 5.2 of AS 2252.1:2010 (Class I) or Section 5.2 of AS 2252.2:2010 (Class II). Testing is needed at least every 12 months and again whenever the cabinet is relocated, has mechanical or electrical work done on it, or has its HEPA filters replaced.