IEC 62304:2015 Medical Device Software Lifecycle Processes
Clause 5: software development process – IEC 62304:2015 Medical Device Software Lifecycle Processes

IEC 62304:2015 Medical Device Software Lifecycle Processes 5.7.3: 5.7.3 Retest after changes

If changes are made during system testing, tests are rerun, adjusted or added to confirm that the change fixes the problem and that it has caused no unintended side effects, and the risk management activities in 7.4 are carried out. Applies to classes A, B and C.