IEC 62304:2015 Medical Device Software Lifecycle Processes
Clause 5: software development process – IEC 62304:2015 Medical Device Software Lifecycle Processes

IEC 62304:2015 Medical Device Software Lifecycle Processes 5.7.4: 5.7.4 Evaluate software system testing

The manufacturer confirms that: the verification strategies and test procedures are suitable; system test procedures can be traced back to the software requirements; every software requirement has been tested or verified in some other way; the links between requirements and tests are recorded; and the test results satisfy the pass or fail criteria. Applies to classes A, B and C.