IEC 60601-1 - Medical Electrical Equipment Safety
Clause 15: construction – IEC 60601-1 - Medical Electrical Equipment Safety

IEC 60601-1 - Medical Electrical Equipment Safety 15.2: Serviceability

What does 15.2 require?

Parts subject to mechanical wear, electrical or environmental degradation or ageing that could create unacceptable risk if left unchecked too long can be reached to inspect, maintain and replace them, and parts likely to be replaced or adjusted are located and secured so that inspection, servicing, replacement and adjustment neither damage nor disturb nearby parts or wiring, checked by inspecting those parts and their location.