IEC 60601-1 - Medical Electrical Equipment Safety
Clause 14: programmable electrical medical systems – IEC 60601-1 - Medical Electrical Equipment Safety

IEC 60601-1 - Medical Electrical Equipment Safety 14.11: 14.11 PEMS validation

A PEMS validation plan covers the validation of basic safety and essential performance and requires checks for unintended functioning; the methods are documented, validation is performed to the plan and its results documented; whoever is overall responsible for validation is independent of the design team, and why that degree of independence was chosen is recorded, no design team member validates their own design, all professional relationships between validation and design team members are documented in the risk management file, and a reference to the validation methods and results is included there.