Australia Therapeutic Goods (Medical Devices) Regulations 2002
Part 10: sponsor identification and custom-made device notices – Australia Therapeutic Goods (Medical Devices) Regulations 2002

Australia Therapeutic Goods (Medical Devices) Regulations 2002 r10.3-10.3A: Custom-made devices: notify the Secretary within 2 months of first manufacture in Australia or first import, and give an annual report on the approved form before 1 October covering all custom-made devices made or imported in the previous financial year

What does r10.3-10.3A require?

A manufacturer of a custom-made device made in Australia must, within 2 months after first making it, give the Secretary its name and business address and a description of the kinds of custom-made devices it makes, with their nomenclature codes; a sponsor importing a custom-made device must do the same within 2 months of first import, adding the overseas manufacturer's details. Each manufacturer and importing sponsor must also give the Secretary, before 1 October each year, a written report on the approved form covering all custom-made devices it made or imported in the previous financial year, including supply details the form asks for. Each failure is an offence (10 penalty units).