Where any planned dealing would release aerosols carrying Risk Group 2 GM micro-organisms, a biological safety cabinet must be present in the laboratory, or some other aerosol containment equipment the Regulator has approved in writing. A Class I or Class II cabinet has to be installed to the requirements of AS 2252.4.
The graph holds this control, the 0 it maps to, and the evidence behind each claim, over MCP and REST.