Australia OGTR Guidelines for Certification of a PC2 Laboratory (Version 3.2) A6: Containment of aerosols

What does A6 require?

Where any planned dealing would release aerosols carrying Risk Group 2 GM micro-organisms, a biological safety cabinet must be present in the laboratory, or some other aerosol containment equipment the Regulator has approved in writing. A Class I or Class II cabinet has to be installed to the requirements of AS 2252.4.